+886-2-8772-8910
Type No. |
L9518 |
L10804 |
L9519 |
L10904 |
L9841 |
Unit |
||
---|---|---|---|---|---|---|---|---|
Type |
Standard |
See-through |
Standard |
See-through |
Standard |
— |
||
Window Material |
UV glass |
Synthetic silica |
MgF2 |
— |
||||
Spectral Distribution |
185 to 400 |
160 to 400 |
115 to 400 |
nm |
||||
Aperture Diameter |
0.5 |
mm |
||||||
|
Drift Max. |
±0.3 |
%/h |
|||||
Output Stability at 230 nm |
Fluctuation (p-p) Typ. |
0.005 |
% |
|||||
Guaranteed Life at 230 nm * |
2000 |
h |
||||||
Discharge Starting Voltage Max. |
400 |
V dc |
||||||
Anode Current |
300 ± 30 |
mA dc |
||||||
Tube Voltage Typ. |
90 |
85 |
V dc |
|||||
Filament Ratings |
Warm-up |
Voltage |
2.5 ± 0.25 |
V dc |
||||
Current Typ. |
4 |
A dc |
||||||
Time Min. |
20 |
s |
||||||
Operating |
Voltage |
1.7 ± 0.2 |
V dc |
|||||
Current Typ. |
3.3 |
A dc |
||||||
Operating Ambient Temperature |
+10 to +50 |
°C |
||||||
Weight |
60 |
65 |
g |
* Lamp life end is defined as the point when light output at 230 nm falls to 50 % of its initial value or when output fluctuation exceeds 0.05 % (p-p).
Parameter |
C9559 |
M9521 |
Unit |
|||||
---|---|---|---|---|---|---|---|---|
Input |
Input voltage |
AC 100 V to AC 240 V |
DC 24 V ± 2.4 V |
— |
||||
Input current (Max.) |
1.4 |
3 |
A |
|||||
Output |
Output voltage (DC) |
With load (Typ.) |
85 / 90 |
V dc |
||||
Without load (Min.) |
200 |
V dc |
||||||
Output current (DC) |
300 ± 30 |
mA dc |
||||||
Current fluctuation (p-p) (Typ.) |
0.005 |
% |
||||||
Current drift at 25 °C (Typ.) |
±0.02 |
%/h |
||||||
Filament ratings |
Warm-up |
Voltage |
2.5 ± 0.2 |
V dc |
||||
Current (Typ.) |
4 |
A dc |
||||||
Operation |
Voltage |
1.7 ± 0.2 |
V dc |
|||||
Current (Typ.) |
3.3 |
A dc |
||||||
Filament warm-up time |
Approx. 20 |
s |
||||||
Trigger voltage |
Anode |
600 |
V peak |
|||||
Auxiliary electrode |
600 |
V peak |
||||||
Cooling method |
— |
Forced air cooling (0.3 m3/min) |
— |
|||||
Operation ambient temperature |
0 to +40 |
°C |
||||||
Storage temperature |
-10 to +60 |
°C |
||||||
Operating and storage humidity |
Below 80 (No condensation) |
% |
||||||
External control (Lamp ON/OFF, Lamp irradiation signal) |
Yes |
Yes |
— |
|||||
Conformance |
CE |
Yes |
Yes |
— |
||||
UL (File No. E249677) |
— |
Yes |
— |